Neptune 2 Waste Management System

Control Panel Improvements

New controls 1
Improved Suction Controls

  • Added LOW, MED, HIGH range color coding (using scale, labels and colors found on wall regulators)
  • Enhanced user control in LOW (50-75 mm-Hg) and MED (80-115 mm-Hg) ranges by rescaling the knobs to allow adjustment at 5 mm-Hg increments

New controls 2

Enhanced Suction Display & Functionality

  • Reduced maximum suction setting from 530 to 480 mm-Hg
  • Added LOW, MED, HIGH range indicator to screen
  • Vacuum units of measure are displayed in mm-Hg only

New controls 3

Added Interactive Alerts

  • Added warning against use with chest tubes and wound drains, which displays on each power up
  • User must read warning and press “OK” before suction can be activated
  • Added audible and visual signals when HIGH (120-480 mm-Hg) suction is selected or activated
  • Display screen colors change from blue to white when HIGH range is selected